ASTM D4327-17

contributor authorASTM - ASTM International
date accessioned2025-09-30T19:25:35Z
date available2025-09-30T19:25:35Z
date copyright2025
date issued2025
identifier otherd4327-17r25.pdf
identifier urihttp://yse.yabesh.ir/std;jsessionid=19EDF98DDCEABA1787A29D9D7C83761E/handle/yse/343472
description abstract5.1 Ion chromatography provides for both qualitative and quantitative determination of seven common anions, F−, Cl−, NO2−, HPO4−2, Br−, NO3−, and SO4−2, in the milligram per litre range from a single analytical operation requiring only a few millilitres of sample and taking approximately 10 min to 15 min for completion. Additional anions, such as carboxylic acids, can also be quantified. Note 2: This test method may be used to determine fluoride if its peak is in the water dip by adding 1 mL of eluent (at 100× the concentration in 8.3) to all 100 mL volumes of samples and standards to negate the effect of the water dip. (See 6.3, and also see 6.4.) The quantitation of unretained peaks should be avoided. Anions such as low molecular weight organic acids (formate, acetate, propionate, etc.) that are conductive coelute with fluoride and would bias fluoride quantitation in some drinking waters and most wastewaters. The water dip can be further minimized if measures are taken to remove carbonic acid which remain in the eluent after suppression using carbonate based eluents. There is no water dip if hydroxide eluents are used. 5.2 Anion combinations such as Cl−/Br− and NO2−/NO3−, which may be difficult to distinguish by other analytical methods, are readily separated by ion chromatography.
languageEnglish
titleStandard Test Method for Anions in Water by Suppressed Ion Chromatographyen
titleASTM D4327-17num
typestandard
statusActive
treeASTM - ASTM International:;2025
contenttypefulltext
scope1.1 This test method2,3 covers the sequential determination of fluoride, chloride, nitrite, ortho-phosphate, bromide, nitrate, and sulfate ions in water by suppressed ion chromatography. Note 1: Order of elution is dependent upon the column used; see Fig. 1.  Sample loop: 50 μL.
identifier DOI10.1520/D4327-17R25