Show simple item record

English -- Medical electrical equipment — Part 2-43: Particular requirements for the basic safety and essential performance of X-ray equipment for interventional procedures - Second Edition;
French -- Appareils électromédicaux — Partie 2-43 : Exigences particulières pour la sécurité de base et les performances essentielles des appareils à rayonnement X lors d’interventions - Second Edition

contributor authorCSA - CSA Group
date accessioned2017-09-04T17:50:02Z
date available2017-09-04T17:50:02Z
date copyright2011.11.01
date issued2011
identifier otherEVQUUEAAAAAAAAAA.pdf
identifier urihttps://yse.yabesh.ir/std/handle/yse/173288
description abstractReplacement:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of X-RAY EQUIPMENT declared by the MANUFACTURER to be suitable for RADIOSCOPICALLY GUIDED INTERVENTIONAL PROCEDURES, hereafter referred to as INTERVENTIONAL X-RAY EQUIPMENT. Its scope excludes, in particular:
– equipment for RADIOTHERAPY;
– equipment for COMPUTED TOMOGRAPHY;
– ACCESSORIES intended to be introduced into the PATIENT;
– mammographic X-RAY EQUIPMENT;
– dental X-RAY EQUIPMENT.
NOTE 1 Examples of RADIOSCOPICALLY GUIDED INTERVENTIONAL PROCEDURES, for which the use of INTERVENTIONAL X-RAY EQUIPMENT complying with this standard is recommended, are given in Annex AA.
NOTE 2 Specific requirements for magnetic navigation devices, and for the use of INTERVENTIONAL X-RAY EQUIPMENT in an operating room environment were not considered in this particular standard; therefore no specific requirements have been developed for these devices or uses. In any case, such devices or uses remain under the general clause requirements.
NOTE 3 INTERVENTIONAL X-RAY EQUIPMENT when used in cross-sectional imaging mode (sometimes described as CT-like mode or cone-beam CT) is covered by this particular standard and not by IEC 60601-2-44 [2]2. Additional requirements for operation in CT-like mode or cone-beam CT were not considered in the present standard.
INTERVENTIONAL X-RAY EQUIPMENT declared by the MANUFACTURER to be suitable for RADIOSCOPICALLY GUIDED INTERVENTIONAL PROCEDURES, which does not include a PATIENT SUPPORT as part of the system, is exempt from the PATIENT SUPPORT provisions of this standard.
If a clause or subclause is specifically intended to be applicable to INTERVENTIONAL X-RAY EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to INTERVENTIONAL X-RAY EQUIPMENT and to ME SYSTEMS, as relevant.
NOTE 4 See also 4.2 of the general standard.
languageEnglish, French
titleCSA CAN/CSA-C22.2 NO. 60601-2-43:11num
titleEnglish -- Medical electrical equipment — Part 2-43: Particular requirements for the basic safety and essential performance of X-ray equipment for interventional procedures - Second Editionen
titleFrench -- Appareils électromédicaux — Partie 2-43 : Exigences particulières pour la sécurité de base et les performances essentielles des appareils à rayonnement X lors d’interventions - Second Editionother
typestandard
page132
statusActive
treeCSA - CSA Group:;2011
contenttypefulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record