UL Standard for Safety Line Isolation Monitors - Fifth Edition

contributor authorUL - UL
date accessioned2017-09-04T18:47:27Z
date available2017-09-04T18:47:27Z
date copyright04/16/2012
date issued2012
identifier otherKLXWUEAAAAAAAAAA.pdf
identifier urihttp://yse.yabesh.ir/std/handle/yse/228601
description abstractThese requirements cover dynamic line isolation monitors and related supplementary indicating units for supervising isolated power-supply circuits in patient care areas of health care facilities, including inhalation-anesthetizing locations in accordance with the National Electrical Code, NFPA 70.
The detectors and supplementary indicating units are intended to be:
a) Included as part of an isolated power-supply center for patient care locations of health care facilities; and
b) Installed in an Other-Than-Hazardous (Classified) Location of a health care facility as defined by the National Electrical Code, NFPA 70.
These requirements do not cover ground fault circuit interrupters for use on grounded alternating current systems in accordance with the National Electrical Code, NFPA 70.